Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242462

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-24

    (66 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for CEREC Cercon 4D™ CAD/CAM Blocks are:

    • Crowns and bridges in the posterior and anterior tooth region
    • Bridges with a maximum of two pontics
    • Inlays, onlays and veneers
    Device Description

    CEREC Cercon 4D™ CAD/CAM Block are pre-sintered zirconia blocks composed of yttria stabilized zirconia, which are designed for CAD/CAM milling. The proposed CEREC Cercon 4D™ CAD/CAM Blocks possess a 3-dimensional contour of dentine powder to mimic the natural tooth build-up. The aim of the 3D contour technology is to achieve more aesthetic restorations than with blocks based on the current multi-layer technology. The CEREC Cercon 4D™ CAD/CAM Blocks are used for fabrication of a dental restoration. The CEREC Cercon 4D™ CAD/CAM Blocks are not provided as the finished, fully finished dental restoration. The blocks are single use, meaning that they can only be milled once and are not recommended for re-milling. The blocks are materials supplied to dental professionals that must be further processed/manufactured using CAD/CAM milling and subsequent sintering to fabricate all ceramic restorations and they are not intended to be reused as in the context of direct patient-applied devices and materials. Therefore, no reprocessing, cleaning or disinfection instructions apply because it is a dental material and not a finished dental restoration itself. After sintering of the restoration, the dental professional will finish the restoration via polishing or glazing, before insertion into the patient's mouth.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "CEREC Cercon 4D CAD/CAM Blocks," a dental material. The information provided is for regulatory clearance and focuses on demonstrating substantial equivalence to existing predicate devices, rather than a clinical effectiveness study of an AI-powered diagnostic device.

    Therefore, many of the requested fields related to AI performance, ground truth, expert adjudication, and comparative effectiveness studies are not applicable or cannot be extracted from this document.

    Here's the information that can be extracted and a clear indication of what is not applicable or not provided in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally "Pass" for conformance to ISO standards and biocompatibility tests. The reported device performance is that it passed these tests.

    Acceptance Criteria (Test Performed)StandardReported Device Performance
    Mechanical/Physical Performance (ISO 6872:2015 Amd 1. 2018)
    Flexural StrengthISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsPass
    CTE (Coefficient of Thermal Expansion)ISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsPass
    SolubilityISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsPass
    RadioactivityISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsPass
    Shrinkage factor (Class 5 Type II material)ISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsPass
    UniformityISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsPass
    Freedom from extraneous materialsISO 6872:2015 Amd 1. 2018 Dentistry Ceramic MaterialsPass
    Compatibility (ISO 9693:2019)
    Debonding / crack-initiationISO 9693:2019 Dentistry - Compatibility testing for metal-ceramic and ceramic-ceramic systemsPass
    Thermal shock resistanceISO 9693:2019 Dentistry - Compatibility testing for metal-ceramic and ceramic-ceramic systemsPass
    Biocompatibility
    ISO 7405: 2018 (Biocompatibility in dentistry)ISO 7405: 2018Meets requirements
    ISO 10993-1:2018 (Biological evaluation)ISO 10993-1:2018Meets requirements
    ISO 10993-5:2009 (Cytotoxicity)ISO 10993-5:2009Meets requirements
    ISO 10993-10:2021 (Skin Sensitization)ISO 10993-10:2021Meets requirements
    ISO 10993-18:2020 (Chemical Characterization)ISO 10993-18:2020Meets requirements
    ISO 10993-23:2021 (Irritation)ISO 10993-23:2021Meets requirements
    USP <151> (Pyrogen Test)USP <151>Meets requirements

    2. Sample size used for the test set and the data provenance

    This document describes non-clinical bench testing and biocompatibility assessments, not a study involving a "test set" of patient data for an AI algorithm. Therefore, this information is not applicable in the context of an AI device. The tests performed are on material samples, and the document does not specify the number of samples used for each test. The provenance of the test samples (e.g., country of origin) is not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is for a dental material, not an AI diagnostic device requiring expert interpretation of a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This document pertains to materials testing, not diagnostic performance with a test set requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is adherence to established international standards (ISO standards) for ceramic dental materials and biocompatibility assessment. The "truth" is whether the material meets the specified physical, mechanical, and biological performance requirements defined by these standards.

    8. The sample size for the training set

    Not applicable. This is not an AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI device that requires a training set.